DIAGnostic:SERVice...    Related Topics

This subsystem provides access to service and diagnostic routines used in service, maintenance and repair. In accordance with the SCPI standard all commands are device-specific.

Service functions are password-protected (SYSTem:PASSword[:CENable]) and should be used by a R&S service representative only. Refer to the service manual for more information.

 

Command tree: DIAGnostic:SERVice

:ALC

:SETTings

[:STATe]

:FUNCtion

:RFPower

:SFUNction


DIAGnostic:ALC:SETTings[:STATe] ON | OFF

Enables or disables the Automatic Level Control for all channels.

<Ch>

Channel number.

 

<Boolean>

ON | OFF - ALC on or off.

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

DIAG:ALC:SETT OFF

Disable the ALC function.  

 

 


DIAGnostic:SERVice:FUNCtion <numeric_value>,...

DIAGnostic:SERVice:SFUNction '<string>',...

Activates a service function (for internal use only).


DIAGnostic:SERVice:RFPower <Boolean>

Turns the internal source power at all ports on or off.   

<Boolean>

ON | OFF - Switches the internal source power on or off.

*RST value

ON

 

SCPI, Command Types

Device-specific, command or query

 

Example:

DIAG:SERV:RFP OFF

Turn off the RF source power.